LVT Test Laboratories
IEC 61326-1 Standard
IEC 61326-1 Standard

TS EN 61326-1 standardı, Elektromanyetik Uyumluluk (EMC) ürün standartlarından biridir. Ölçme, kontrol ve laboratuvar ürün üreticileri, ürünlerini TS EN 61326-1 standardına göre tasarlar ve test (EMC

IEC 61326-1 Standard

IEC 61326-1 Standard

IEC 61326-1 Standard

ELECTRICAL EQUIPMENT FOR MEASUREMENT, CONTROL AND LABORATORY USE - EMC REQUIREMENTS - Part 1: GENERAL REQUIREMENTS

TS EN 61326-1 standard is one of the Electromagnetic Compatibility (EMC) product standards. Measurement, control and laboratory product manufacturers design their products according to TS EN 61326-1 standard and have them tested (EMC Test). 

TS EN 61326-1 standard; Electrical equipment for measurement, control and laboratory covers the immunity and emission requirements regarding electromagnetic compatibility (EMC) for electrical equipment operating from an electrical grid or a battery. 

The purpose of the TS EN 61326-1 standard is to specify the general characteristics and experiments (EMC Test) for electrical equipment electromagnetic compatibility (EMC) for measurement, control and laboratory.


Electromagnetic Compatibility (EMC) Tests:

Electromagnetic Compatibility Tests (EMC Test) are aimed to operate any electrical and electronic device or system without being affected by and affecting other devices or systems in the environment. EMC Tests are examined in 2 general categories as immunity and emission.

Emission Tests specified in 61326-1;

  • Radiated Emission Test - CISPR 11 / EN 55011
  • Conducted Emission Test - CISPR 11 / EN 55011
  • Harmonics Test - EN 61000-3-2
  • Flicker Test - EN 61000-3-3

Immunity Tests specified in 61326-1;

  • Electrostatic Discharge (ESD) Test - EN 61000-4-2
  • Radiated Immunity Test - EN 61000-4-3
  • Electrical Fast Transients / Burst Test - EN 61000-4-4
  • Surge Testi - EN 61000-4-5
  • Conducted Immunity Test - EN 61000-4-6
  • Magnetics Test - EN 61000-4-8
  • Voltage Dips and Interruptions Test - EN 61000-4-11

Immunity Test Specifications


Immunity test requirements for equipment  intended to be used in a BASIC ELECTROMAGNETIC ENVIRONMENT

Port

EMC Test

Immunity Test Value





Enclosure

ESD Test

EN 61000-4-2

Contact: ±4 kV

Air:±8 kV

Radiated Immunity Test EN 61000-4-3

80 MHz – 1 GHz (3 V/m)

1,4 GHz – 6 GHz (3 V/m)


%80 AM – 1 kHz

Magnetic Fields Test

EN 61000-4-8

3 A/m

50-60 Hz







AC Power

Voltage Dips and Interruptions Test -  EN 61000-4-11

During half cycle % 0

During 1 cycle % 0

During 25/30 cycles % 70

During 250/300 cycles % 0

Electrical Fast Transients / Burst Test

 EN 61000-4-4

1 kV (5/50 ns, 5 or 100kHz)

Surge Test - EN 61000-4-5

Line to line;0,5 kV

Line to ground;1 kV

Conducted Immunity Test - EN 61000-4-6

3 V (150 kHz to 80 MHz)




DC Power

Electrical Fast Transients / Burst Test

EN 61000-4-4

1 kV (5/50 ns, 5 or 100kHz)

Surge Test - EN 61000-4-5

Line to line;0,5 kV

Line to ground;1 kV

Conducted Immunity Test - EN 61000-4-6

3 V (150 kHz to 80 MHz)


Immunity test requirements for equipment intended to be used in an INDUSTRIAL ELECTROMAGNETIC ENVIRONMENT 


Port

EMC Test

Immunity Test Value





Enclosure

ESD Test

EN 61000-4-2


Contact: ±4 kV

Air:±8 kV

Radiated Immunity Test EN 61000-4-3

80 MHz – 1 GHz (10 V/m)

1,4 GHz – 6 GHz (3 V/m)

%80 AM – 1 kHz

Magnetic Fields Test

EN 61000-4-8


30 A/m

50-60 Hz







AC Power

Voltage Dips and Interruptions Test -  EN 61000-4-11

During 1 cycle % 0

During 10/12 cycles % 40

During 25/30 cycles % 70

During 250/300 cycles % 0


Electrical Fast Transients / Burst Test

EN 61000-4-4

2 kV (5/50 ns, 5 or 100kHz)

Surge Test - EN 61000-4-5


Line to line;1 kV

Line to ground;2 kV

Conducted Immunity Test - EN 61000-4-6

3 V (150 kHz to 80 MHz)




DC Power

Electrical Fast Transients / Burst Test

EN 61000-4-4

2 kV (5/50 ns, 5 or 100kHz)

Surge Test - EN 61000-4-5


Line to line;1 kV

Line to ground;2 kV

Conducted Immunity Test - EN 61000-4-6

3 V (150 kHz to 80 MHz)


Immunity test requirements for equipment intended to be used in a CONTROLLED ELECTROMAGNETIC ENVIRONMENT 


Port

EMC Test

Immunity Test Value





Enclosure

ESD Test

EN 61000-4-2


Contact: ±4 kV

Air:±8 kV

Radiated Immunity Test EN 61000-4-3

80 MHz – 1 GHz (1 V/m)

1,4 GHz – 6 GHz (1 V/m)

%80 AM – 1 kHz







AC Power

Electrical Fast Transients / Burst Test

EN 61000-4-4

1 kV (5/50 ns, 5 or 100kHz)

Conducted Immunity Test - EN 61000-4-6

1 V (150 kHz to 80 MHz)




DC Power

Elektriksel Hızlı Geçici Rejim / Ani Darbe (Electrical Fast Transients / Burst) Testi

TS EN 61000-4-4

2-1 kV (5/50 ns, 5 or 100kHz)

Conducted Immunity Test - EN 61000-4-6

1 V (150 kHz to 80 MHz)



 Performance criteria 

Performance criterion A: During and after the test the EUT shall continue to operate as intended without operator intervention. No degradation of performance or loss of function is allowed below a minimum performance level specified by the manufacturer when the EUT is used as intended. The performance level may be replaced by a permissible loss of performance. If the minimum performance level or the permissible performance loss is not specified by the manufacturer, then either of these may be derived from the product description and documentation, and by what the user may reasonably expect from the EUT if used as intended.

Performance criterion B: After the test, the EUT shall continue to operate as intended without operator intervention. No degradation of performance or loss of function is allowed, after the application of the phenomena below a performance level specified by the manufacturer, when the EUT is used as intended. The performance level may be replaced by a permissible loss of performance. During the test, degradation of performance is allowed. However, no change of operating state or stored data is allowed to persist after the test. If the minimum performance level (or the permissible performance loss) is not specified by the manufacturer, then either of these may be derived from the product description and documentation, and by what the user may reasonably expect from the EUT if used as intended.

Performance criterion C: During and after testing, a temporary loss of function is allowed, provided the function is selfrecoverable, or can be restored by the operation of the controls or cycling of the power to the EUT by the user in accordance with the manufacturer’s instructions.

See These Topics
5165